A modern, rapid RP-HPLC-UV method was developed and validated in
compliance with FDA and EMA guidelines for simultaneous quantification of 15 βlactam antibiotics) Ampicillin, Amoxicillin, cephalexin, cefotaxime, cefoxitin,
cefamandole, cephalothin, piperacillin, penicillin, oxacillin, cloxacillin, nafcillin
Carbenicillin, Mezlocillin and Dicloxacillin) in pharmaceutical formulations and
pure forms. The method employs Column NEUCLEODUR C-18 (4.0 mm x 100
mm, 5µm particle size), at a temperature of thirty degrees Celsius, and the mobile
phase was acetonitrile and KH2PO4 using gradient elution with a total separation
time of 13 minutes, a flow rate of 1.3 ml/min, at = pH 4.5 for the buffer solution and
the λ max was 220 nm. The met
...